How to measure Gardner color value?

Chemical Industry | 2017-12-09 13:01:49

We receive the liquid sample from one of our customer. They ask us to measure its Gardner color value. Today we will introduce how to measure it with the benchtop Garder color spectrophotometer CS-810.


1. Gardner color spectrophotometer CS-810 Calibration
Black calibrate it with black tile and then click black calibration on software.

Gardner color spectrophotometer 1Gardner color spectrophotometer 2


Gardner color spectrophotometer White Calibration

White calibrate it by distill water with glass cell.

How to measure Gardner color value?Gardner color spectrophotometer 3


Gardner color spectrophotometer Air Calibration

Air calibrate the instrument. (Note: make sure nothing is on the test aperture and then press “Air” on software for air calibration.)

Gardner color spectrophotometer 4


2. Prepare the samples.

Gardner color spectrophotometer 5


3. Choose “Gardner” and “light source” and “angle” on software.
“Setup”—“list option”---“index”---“Gardner Color”
“Setup”—“Angle/Source Setting”—“C and 2 degree”
Gardner color spectrophotometer 6Gardner color spectrophotometer 7

4. Put the samples with glass cell into the Gardner color spectrophotometer one by one. Press “target” and then “sample” on software. We get the Gardner color value directly on the PC software.
Gardner color spectrophotometer 8Gardner color spectrophotometer 9

If any inquiry about liquid or solid color measurement, please contact us freely.